Remote control commands
R&S
®
RTP
1977User Manual 1337.9952.02 ─ 12
BUS<m>:EBTB:DPLus:SOURce <SourceDplus>
Selects the D+ source of the provided single-ended signal, if BUS<m>:EBTB:TYPE
SINGle is set.
Suffix:
<m>
.
1..4
Parameters:
<SourceDplus> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4
*RST: C1W1
Usage: Asynchronous command
BUS<m>:EBTB:DPLus:THReshold
<ThresholdDplus>
Sets the high threshold (D +) of the signal, if BUS<m>:EBTB:TYPE SINGle.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdDplus> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
BUS<m>:EBTB:FCSY <CustSync>
Sets a pattern value, if BUS<m>:EBTB:SYNC CUS.
Suffix:
<m>
.
1..4
Parameters:
<CustSync>
BUS<m>:EBTB:SCOunt?
Returns the symbol count for the selected serial bus, i.e. the number of symbols in the
present acquisition.
Suffix:
<m>
.
1..4
Return values:
<Count> Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only