EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH - Page 448

Rohde & Schwarz Scope Rider RTH
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
448User Manual 1326.1578.02 ─ 12
TRIGger:UART:DATA <Pattern>
Defines the data pattern as trigger condition. Enter the words in MSB first bit order.
Parameters:
<Pattern> String with max. 8 bit in binary format. Characters 0, 1, and X
are allowed.
Example:
TRIGger:UART:DATA '1x11'
TRIGger:UART:DATA?
1X11
TRIGger:UART:DCONdition <Relation>
Defines how the specified data pattern is compared with the acquired signal. The
instrument triggers if the acquired data is equal or unequal to the pattern.
Parameters:
<Relation> EQUal | NEQual
*RST: EQUal
15.11.4.3 UART Decode Results
The frame suffix <m> selects the frame index for which the result is queried.
The byte suffix <n> selects the byte index for which the result is queried.
BUS:UART:WCOunt?..................................................................................................... 448
BUS:UART:WORD<m>:STATe?.......................................................................................448
BUS:UART:WORD<m>:STARt?.......................................................................................449
BUS:UART:WORD<m>:STOP?....................................................................................... 449
BUS:UART:WORD<m>:VALue?...................................................................................... 450
BUS:UART:WCOunt?
Returns the number of decoded symbols (words).
Return values:
<Count> Total number of decoded words.
Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS:UART:WORD<m>:STATe?
Returns the status of the specified symbol (word).