Waveform Setup
R&S
®
Scope Rider RTH
51User Manual 1326.1578.02 ─ 12
► To start and stop acquisition, press the [RUN STOP] key.
3.6.1 General Trigger Settings
General trigger settings are the settings that are independent of the trigger type. The
settings specific for a trigger type are described in the following sections.
Description of settings
Trigger Mode
The trigger mode determines the behavior of the instrument if no trigger occurs, and
also the number of acquired waveforms when a trigger occurs.
"Auto"
The instrument triggers repeatedly after a time interval if the trigger
conditions are not fulfilled. If a real trigger occurs, it takes prece-
dence. This mode helps to see the waveform even before the trigger
is set. Successive waveforms are not triggered at the same point of
the waveform.
"Normal"
The instrument acquires waveforms continuously, each time when a
trigger occurs. If no trigger occurs, no waveform is acquired and the
last acquired waveform is displayed. If no waveform was captured
before, nothing is displayed.
"Single"
When a trigger occurs, the instrument acquires one waveform and
stops the acquisition.
Remote command:
TRIGger:MODE on page 309
Trigger Type
Selects the trigger type, the event type that defines the trigger point.
●
Chapter 3.6.2, "Edge Trigger", on page 53
●
Chapter 3.6.3, "Glitch Trigger", on page 54
●
Chapter 3.6.4, "Width Trigger", on page 55
Trigger