Waveform Setup
R&S
®
Scope Rider RTH
52User Manual 1326.1578.02 ─ 12
Chapter 3.6.5, "Video/TV Trigger", on page 57
R&S RTH-K19 trigger options
Chapter 3.6.7, "Pattern Trigger (R&S RTH-K19)", on page 62
Chapter 3.6.8, "State Trigger (R&S RTH-K19)", on page 64
Chapter 3.6.9, "Runt Trigger (R&S RTH-K19)", on page 65
Chapter 3.6.10, "Slew Rate Trigger (R&S RTH-K19)", on page 67
Chapter 3.6.11, "Data2Clock Trigger (R&S RTH-K19)", on page 68
Chapter 3.6.12, "Serial Pattern Trigger (R&S RTH-K19)", on page 70
Chapter 3.6.13, "Timeout Trigger (R&S RTH-K19)", on page 73
Chapter 3.6.14, "Interval Trigger (R&S RTH-K19)", on page 74
Chapter 3.6.15, "Window Trigger (R&S RTH-K19)", on page 75
Options containing special triggers
Chapter 3.6.16, "Protocol Trigger (R&S RTH-K1, -K2, -K3, -K9 and -K10)",
on page 76
Remote command:
TRIGger:TYPE on page 310
Source
Selects the trigger source, the channel on which the trigger condition is checked. All
possible channels are listed. You can trigger on any channel to which a signal is con-
nected, even if the channel is not active.
For most trigger types, analog and digital channels can be used as trigger source. Digi-
tal channels require option R&S RTH-B1. For video, runt and slew rate trigger, only
analog channels are available.
Remote command:
TRIGger:SOURce on page 310
Trigger Level
Sets the trigger voltage level.
For the Video/TV trigger, the trigger level is the threshold of the sync pulse. Make sure
that the trigger level crosses the synchronizing pulses of the video signal.
Remote command:
TRIGger:LEVel<m>:VALue on page 310
Holdoff Mode
Selects the method to define the holdoff.
The trigger holdoff defines when the next trigger after the current will be recognized.
Thus, it affects the next trigger to occur after the current one. Holdoff helps to obtain
stable triggering when the oscilloscope is triggering on undesired events.
Trigger