Preface
R&S
®
Scope Rider RTH
15Getting Started 1326.1561.02 ─ 10
Input channels
R&S RTH1002: 2 oscilloscope channels, 1 multimeter
R&S RTH1004: 4 oscilloscope channels
Maximum input voltage
At BNC inputs: CAT IV 300 V
With probe R&S RT-ZI10 or R&S RT-ZI11: CAT IV 600 V, CAT III 1000 V
Meter input: CAT IV 600 V; CAT III 1000 V
Voltage ratings: V RMS (50 to 60 Hz) for AC sine wave and V DC for DC applica-
tions.
Vertical system
Analog bandwidth (–3 dB) of oscilloscope channels:
R&S RTH1002 and R&S RTH1004: ≥ 60 MHz
R&S RTH1002 with -B221 option and R&S RTH1004 with -B241 option:
100 MHz
R&S RTH1002 with -B222 option and R&S RTH1004 with -B242 option:
200 MHz
R&S RTH1002 with -B223 option and R&S RTH1004 with -B243 option:
350 MHz
R&S RTH1002 with -B224 option and R&S RTH1004 with -B244 option:
500 MHz
Rise time (calculated):
R&S RTH1002 and R&S RTH1004 with bandwidth of 60 MHz, 100 MHz,
200 MHz, 350 MHz, 500 MHz: < 5.8 ns, < 3.5 ns, < 1.75 ns, < 1 ns, < 700 ps
respectively
DC gain accuracy:
Offset and position set to zero after self-alignment
Input sensitivity > 5 mV/div: ±1.5 % of reading
Input sensitivity > 2 mV/div to 5 mV/div: ±2 % of reading
input sensitivity 2 mV/div: ±2.5 % of reading
Input impedance: 1 M ± 1 % || 12 pF ± 2 pF (meas.)
Input coupling: DC, AC
Input sensitivity: 2 mV/div to 100 V/div
Main specifications