EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 236

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol Analysis
R&S
®
Scope Rider RTH
236User Manual 1326.1578.02 ─ 12
Identifer Relation
Defines how the specified identifier pattern is compared with the acquired signal. The
instrument triggers if the acquired address is equal or unequal the defined pattern.
Remote command:
TRIGger:SENT:ICONdition on page 479
Sync Pulse Error
Detects a synchronization/calibration pulse error in transmission sequences of the fast
channel. An error occurs when:
The duration of the synchronization/calibration pulse (in ticks) is less than 56*(1-
clock tolerance) or more than 56*(1+clock tolerance).
The synchronization/calibration pulse duration of frame (n–1) varies by more than
1.5625% from the calibration/sync pulse duration of frame (n).
Remote command:
TRIGger:SENT:PULSeerror on page 480
Pulse Period Error
Detects an error in the calibration/sync pulse in transmission sequences of the fast
channel. An error occurs when a nibble has any of the following:
Number of ticks at low is fewer than 4 ticks.
Nibble value< 0 (fewer than 12 ticks) or > 15 (more than 27 ticks).
Through the pulse period error, you can also identify a nibble length error of the previ-
ous transmission sequence.
Remote command:
TRIGger:SENT:PPERioderror on page 480
Fast CRC Error
Detects a checksum error in the transmission sequences of the fast channel. The CRC
length is 4 bits.
Remote command:
TRIGger:SENT:FCRCerror on page 479
Slow CRC Error
Detects a checksum error in serial messages of the slow channel. The CRC length is 4
bits for short serial messages, and 6 bit of enhanced serial messages.
Remote command:
TRIGger:SENT:SCRCerror on page 480
Frame Length Error
Detects frame length errors in transmission sequences when Pause Pulse is set to
"Constant frame length".
A frame length error occurs, when the total length of the transmission sequence
(including pause pulse) does not match the Frame Length.
Remote command:
TRIGger:SENT:IRFLength on page 480