Remote Control Commands
R&S
®
Scope Rider RTH
340User Manual 1326.1578.02 ─ 12
CURSor:Y2AMplitude?................................................................................................... 341
CURSor:MEASurement<m>:RESult:ACTual?................................................................... 341
CURSor:MEASurement<m>:RESult:LIMit?....................................................................... 341
CURSor:TDELta?
Returns the time difference Δ between to vertical cursor lines.
Return values:
<ResultDeltaT> Range: -100E+24 to 100E+24
*RST: 0
Usage: Query only
CURSor:ITDelta?
Returns the inverse value of time difference between to vertical cursor lines1/Δt.
Return values:
<ResultDeltaTInv> Range: -100E+24 to 100E+24
Increment: 0
*RST: 0
Usage: Query only
CURSor:X1Position <UserX1>
CURSor:X2Position <UserX2>
Set the horizontal positions t1 and t2 (time) of the vertical cursor lines.
Parameters:
<UserX1>, <UserX2> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
CURSor:DELTa?
Returns the absolute value of the difference between the positions of horizontal cursor
lines Δy.
Return values:
<ResultDelta> Range: -100E+24 to 100E+24
*RST: 0
Usage: Query only
CURSor:Y1Position <UserY1>
CURSor:Y2Position <UserY2>
Set the vertical positions y1 and y2 of the horizontal cursor lines.