EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 363

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
363User Manual 1326.1578.02 ─ 12
Parameters:
<TraceSource> SPECtrum | MAXimum | MINimum | AVERage
*RST: SPECtrum
SPECtrum:MARKer:SETup:DISTance <Distance>
Defines a minimum distance between two frequencies that must be exceeded in order
to detect individual peaks.
Parameters:
<Distance> Range: 0 to 500E+6
Increment: 10
*RST: 1E+6
Default unit: Hz
SPECtrum:MARKer:SETup:EXCursion
<Excursion>
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered a peak. To avoid identifying noise peaks, enter a peak excur-
sion value that is higher than the noise levels.
See also "Peak excursion" on page 125.
Parameters:
<Excursion> Range: 0 to 260
Increment: 1
*RST: 10
Default unit: dB
SPECtrum:MARKer:SETup:MLEVel <Threshold>
Defines an absolute level threshold as an additional condition for the peak search.
Only peaks that exceed the threshold are detected.
Parameters:
<Threshold> Range: -260 to 260
Increment: 1
*RST: -40
Default unit: dB
SPECtrum:MARKer:RCOunt?
Queries the number of markers for which peaks were actually detected during the peak
search. Note that the number of markers is restricted by the SPECtrum:MARKer:
COUNt command.
Return values:
<NumberOfResults> Range: 0 to 15
Increment: 1
*RST: 0
Spectrum Analysis

Table of Contents

Other manuals for Rohde & Schwarz Scope Rider RTH1004

Related product manuals