EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 364

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
364User Manual 1326.1578.02 ─ 12
Usage: Query only
SPECtrum:MARKer:RESult<m>:FREQuency[:VALue]?
Queries the frequency of the specified marker.
Suffix:
<m>
.
1 to 15
Number of the marker. The maximum number depends on how
many markers were selected and how many peaks were detec-
ted (see SPECtrum:MARKer:RCOunt? on page 363).
Return values:
<Frequency> Range: -10E+9 to 10E+9
Increment: 10
*RST: 10E+6
Default unit: Hz
Usage: Query only
SPECtrum:MARKer:RESult<m>:FREQuency:DELTa?
Queries the frequency difference between the reference marker and the specified
marker.
Suffix:
<m>
.
1 to 15
Marker; maximum depends on active markers and detected
peaks (see SPECtrum:MARKer:COUNt and SPECtrum:
MARKer:RCOunt?
Return values:
<FrequencyDelta> Range: -10E+9 to 10E+9
Increment: 10
*RST: 10E+6
Default unit: Hz
Usage: Query only
SPECtrum:MARKer:RESult<m>:LEVel[:VALue]?
Queries the power level of the specified marker in relation to the reference marker (see
SPECtrum:MARKer:RMARker:VALue? on page 365).
Suffix:
<m>
.
1 to 15
Number of the marker. The maximum number depends on how
many markers were selected and how many peaks were detec-
ted (see SPECtrum:MARKer:RCOunt? on page 363).