Remote Control Commands
R&S
®
Scope Rider RTH
373User Manual 1326.1578.02 ─ 12
Usage: Query only
HARMonic:RESult<m>:FUNDamental?
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Return values:
<Frequency> Range: 9 to 1005
Increment: 10E-6
*RST: 0
Default unit: Hz
Usage: Query only
HARMonic:RESult<m>:HARMonics<n>:ABSMagnitude:MAXimum?
HARMonic:RESult<m>:HARMonics<n>:ABSMagnitude:MINimum?
HARMonic:RESult<m>:HARMonics<n>:ABSMagnitude[:CURRent]?
Returns the currently measured, minimum or maximum power level of the selected
harmonic.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Return values:
<AbsMagnitude>
Usage: Query only
HARMonic:RESult<m>:HARMonics<n>:LIMCheck?
Queries whether a limit is defined for the specified harmonic (either by the EN50160
standard or by a user-defined configuration file). Only if a limit value is defined, a limit
check is performed for the harmonic.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Return values:
<State> ON | OFF
ON
A limit is specified and a limit check will be performed.