Remote Control Commands
R&S
®
Scope Rider RTH
380User Manual 1326.1578.02 ─ 12
The METer:CONFigure:<function> commands are only for configuration. To acti-
vate a measurement, use METer:SENSe:FUNCtion.
METer:CONFigure:VALue? returns the active measurement and range.
To read the result, use METer<m>:READ? or METer<m>:FETCh?.
METer:CONFigure:CONTinuity........................................................................................ 380
METer:CONFigure:DIODe...............................................................................................380
METer:CONFigure:TEMPerature..................................................................................... 380
METer:CONFigure:CAPacitance...................................................................................... 380
METer:CONFigure:CURRent:AC..................................................................................... 380
METer:CONFigure:CURRent:DC..................................................................................... 380
METer:CONFigure:FREQuency....................................................................................... 381
METer:CONFigure:RESistance........................................................................................381
METer:CONFigure:VOLTage:AC...................................................................................... 381
METer:CONFigure:VOLTage:DC......................................................................................381
METer:CONFigure:CONTinuity
METer:CONFigure:DIODe
METer:CONFigure:TEMPerature
Configures the specified measurement. The instrument sets a fixed range.
Usage:
Event
METer:CONFigure:CAPacitance
Sets the internal parameters and configures the range for capacitance measurements.
Parameters:
<Range> <numeric value> | 'AUTO' | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 10 nF|100 nF|
1 μF|10 μF|100 μF|1 mF|10 mF.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range: 10 nF to 10 mF
*RST: 10 nF
Usage: Setting only
METer:CONFigure:CURRent:AC
METer:CONFigure:CURRent:DC
Sets the internal parameters and configures the range for current measurements.
You need an external shunt resistor or I/U converter for current measurement.
Digital Multimeter (R&S
RTH1002)