Remote Control Commands
R&S
®
Scope Rider RTH
412User Manual 1326.1578.02 ─ 12
*RST: INTernal
COUNter<m>:SENSe:REFerence:VALue <RefFreq>
Defines the nominal frequency value to be used as a reference for counter 2.
Suffix:
<m>
.
1..2
Parameters:
<RefFreq> Range: 10 to 500E+6
Increment: 1
*RST: 10E+6
Default unit: Hz
COUNter<m>:SENSe:SOURce
<InputChannel>
Selects the channel used for input for the selected counter.
Suffix:
<m>
.
1..2
Parameters:
<InputChannel> C1 | C2 | C3 | C4
*RST: C1
COUNter<m>:SENSe:STATe <State>
Activates or deactivates the counter.
Suffix:
<m>
.
1..2
Parameters:
<State> ON | OFF
*RST: OFF
COUNter<m>:SENSe:TRIGger:MODE <TriggerMode>
Determines the measurement mode.
Suffix:
<m>
.
1..2
Parameters:
<TriggerMode> AUTO | SINGle
AUTO
Continuous measurement; the most recent value is preserved