Remote Control Commands
R&S
®
Scope Rider RTH
424User Manual 1326.1578.02 ─ 12
LOGGer:MEASurement<m>:ENABled?
Returns the measurement state of scope and meter measurements.
Suffix:
<m>
.
1..4
Return values:
<Enabled> ON | OFF
*RST: OFF
Usage: Query only
LOGGer:MEASurement<m>:TYPE?
Returns the measurement type of the selected measurement.
Suffix:
<m>
.
1..4
Return values:
<Type> See MEASurement<m>:TYPE on page 336.
Usage: Query only
LOGGer:MEASurement<m>:SOURce?
Returns the source channel of the selected measurement.
Suffix:
<m>
.
1..4
Return values:
<Source> C1 | C2 | C3 | C4 | M1 | R1 | XY | D0 | D1 | D2 | D3 | D4 | D5 |
D6 | D7
<Source2> C1 | C2 | C3 | C4 | M1 | R1 | XY | D0 | D1 | D2 | D3 | D4 | D5 |
D6 | D7
Example:
LOGGer:MEASurement4:SOURce?
<-- C2,C1
Measurement 4 is a delay measurement on source 1 = C2 and
source 2 = C1
Usage: Query only
15.10.6 Logger Records
You can export the data of an active slot to a file, in CSV or MAT format. See also:
Chapter 8.6, "Export of Logger Records", on page 166.
LOGGer:SLOT:ACTive?..................................................................................................425
LOGGer:SLOT:EXPort:NAME..........................................................................................425
LOGGer:SLOT:EXPort:SAVE...........................................................................................425
Data Logging

Table of Contents

Other manuals for Rohde & Schwarz Scope Rider RTH1004

Related product manuals