EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 438

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
438User Manual 1326.1578.02 ─ 12
CMOS
2.5 V
USER
Set a user-defined value for each line using the
BUS:SPI:<line>:THReshold commands.
*RST: TTL
Usage: SCPI confirmed
BUS:SPI:SCLK:THReshold
<ThresholdValue>
BUS:SPI:SSEL:THReshold <ThresholdValue>
BUS:SPI:MOSI:THReshold <ThresholdValue>
BUS:SPI:MISO:THReshold <ThresholdValue>
Set a user-defined value for the corresponding line if BUS:SPI:TECHnology is set to
USER.
Parameters:
<ThresholdValue> Range: -400 to 400
Increment: 1E-3
*RST: 1.4
Default unit: V
BUS:SETReflevels
Sets the appropriate threshold or threshold for the selected serial protocol.
Usage:
Event
BUS:SPI:WSIZe <WordLength>
Sets the number of bits in a word.
Parameters:
<WordLength> WL4Bit | WL8Bit | WL12bit | WL16bit | WL20bit | WL24bit |
WL28bit | WL32bit
*RST: WL8Bit
BUS:SPI:ORDer <BitOrder>
Defines if the data of the words starts with MSB (most significant bit) or LSB (least sig-
nificant bit). Results are displayed in the specified order.
Parameters:
<BitOrder> LSBF | MSBF
*RST: MSBF