EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 44

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Waveform Setup
R&S
®
Scope Rider RTH
44User Manual 1326.1578.02 ─ 12
Remote command:
CHANnel<m>:SCALe on page 300
CHANnel<m>:RANGe on page 300
FFT mode:
SPECtrum:FREQuency:MAGNitude:SCALe on page 353
Spectrum mode:
SPECtrum:FREQuency:SCALe on page 358
Counter mode:
COUNter<m>:SENSe:RANGe on page 411
[POS] keys
Move the selected signal up or down in the diagram. The position is a graphical setting
given in divisions, while the offset sets a voltage.
You can also drag the channel marker on the screen.
Remote command:
CHANnel<m>:POSition on page 302
Spectrum mode:
SPECtrum:FREQuency:POSition on page 357
3.2.2 Effect of the Bandwidth Filter
Low-pass filters reduce the speed of the signal inside the instrument and cause a delay
of the signal on the screen. The delay time depends on the selected filter.
The following table lists the approximate delay of the signal that is caused by various
filters.
Table 3-1: Approximate signal delay dependent on the bandwidth filter
Filter Approx. delay
Filter Approx. delay
200 MHz 30.2 ns
500 kHz 9.07 μs
100 MHz 30.7 ns
200 kHz 22.13 μs
50 MHz 138.5 ns
100 kHz 43.87 μs
20 MHz 145 ns
50 kHz 87.47 μs
10 MHz 166.5 ns
20 kHz 218 μs
5 MHz 193 ns
10 kHz 434.7 μs
2 MHz 270.5 ns
5 kHz 869.3 μs
1 MHz 4.71 μs
2 kHz 2.173 ms
1 kHz 4.347 ms
Vertical Setup

Table of Contents

Other manuals for Rohde & Schwarz Scope Rider RTH1004

Related product manuals