EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 484

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
484User Manual 1326.1578.02 ─ 12
Return values:
<StatusCommunication>Range: 0 to 0
Increment: 1
*RST: 0
Usage: Query only
BUS:SENT:FRAMe<m>:SDATa?
Returns the symbolic data of the frame.
Suffix:
<m>
.
*
Return values:
<SymbolicData> Comma-separated list of values. The first value is the number of
bytes, followed by the decoded data bytes.
Usage: Query only
BUS:SENT:FRAMe<m>:STARt?
Returns the start time of the selected frame.
Suffix:
<m>
.
*
Return values:
<FrameStart> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS:SENT:FRAMe<m>:STATus?
Returns the overall state of the selected frame.
Suffix:
<m>
.
*
Return values:
<FrameState> OK | SYNC | PULSe | CRC | IRFL | FORM | INSufficient
OK: The frame is valid.
SYNC: Synchronization error occurred.
PULse: Pulse error occurred.
CRC: Cyclic redundancy check failed.
IRFL: Irregular frame length error occurred.
FORM: Format error occurred.
INSufficient: The frame is not completely contained in the
acquisition. The acquired part of the frame is valid.
Protocol Analysis