Waveform Setup
R&S
®
Scope Rider RTH
73User Manual 1326.1578.02 ─ 12
See also Chapter 3.6.12.1, "Pattern Definition", on page 70.
Remote command:
TRIGger:SPATtern:PATTern on page 329
Bit Order
Defines if the data words start with MSB (most significant bit) or LSB (least significant
bit).
Remote command:
TRIGger:SPATtern:ORDer on page 329
3.6.13 Timeout Trigger (R&S RTH-K19)
The timeout trigger checks if the signal stays above or below the trigger level for a
specified time lapse. In other words, the trigger occurs if the signal does not cross the
trigger level during the specified time.
[SETUP] key > "Trigger Type" = "Timeout"
Description of settings
Figure 3-19: Timeout trigger
Range
Selects the relation of the signal level to the trigger level:
"Stays High"
The signal level stays above the trigger level.
"Stays Low"
The signal level stays below the trigger level.
"High or Low"
The signal level stays above or below the trigger level.
Remote command:
TRIGger:TIMeout:RANGe on page 330
Time
Defines the time limit for the timeout at which the instrument triggers.
Remote command:
TRIGger:TIMeout:TIME on page 330
Trigger