Signal Generator Settings
R&S
®
SGS100A
103User Manual 1173.9105.02 ─ 12
Using USB hub
Figure 7-2 illustrates the connection as principle. For details, see the description
R&S
®
NRP
®
Series Power Sensors getting started.
Figure 7-2: Connecting power sensors to the R&S
SGS via USB hub
I, II = Connection order
G = Signal source (R&S SGS)
1 = USB type Micro-A connector
2 = USB Adapter Micro-A to A
3 = Standard USB cable with USB type A and USB type B connectors
4 = USB type B connector
4a = External power supply unit and extra power cable
5 = Cable R&S NRP-ZKU per sensor of the R&S NRPxx power sensor family; R&S NRP-Z3 or R&S NRP-
Z4 USB adapter cables for sensors of the R&S NRP-Zxx family
Connection order
Always connect the equipment in the following order:
1. Connect the R&S NRP-Z5 sensor hub or the USB hub to the power supply and to
the R&S SGS
2. Switch on the R&S SGS
3. Connect/disconnect the R&S NRP sensors
Detection and mapping
The R&S SGS automatically detects a connected R&S NRP power sensor and indi-
cates it in the "NRP Power Viewer" and "NRP Sensor Mapping" dialogs.
You can change the default mapping in the NRP Sensor Mapping dialog.
On connection, the R&S SGS immediately starts the measurement of a detected
R&S NRP power sensor. If you perform an instrument preset, the R&S SGS stops the
measurements. The connection and the mapping of the power sensors remain, the
measurements must be restarted.
7.5.8.2 NRP Sensor Mapping
The "NRP Sensor Mapping" lists all R&S NRP sensors detected by the instrument.
Level and Power-On Settings