Signal Generator Settings
R&S
®
SGS100A
115User Manual 1173.9105.02 ─ 12
7.6.2 Pulse Generator Settings
► To access the pulse connector/trigger settings, select "SGMA-GUI > Instrument
Name > Pulse Modulation > Pulse Genrator".
Comprises the settings necessary to configure the internal pulse modulation signal.
Pulse Mode
Sets the operating mode of the pulse generator.
"Single"
Generates a single pulse in one pulse period, see Figure 7-3.
"Double"
Generates two pulses in one pulse period, see Figure 7-4.
Remote command:
[:SOURce<hw>]:PULM:MODE on page 256
Pulse Period
Sets the repetition rate of the generated pulse signal, see "Pulse modulation signal
waveforms" on page 113.
Remote command:
[:SOURce<hw>]:PULM:PERiod on page 256
Pulse Width
Sets the pulse duration of the generated pulse signal, see "Pulse modulation signal
waveforms" on page 113.
Remote command:
[:SOURce<hw>]:PULM:WIDTh on page 259
Double Pulse Width
Sets the width of the second pulse, see Figure 7-4.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:WIDTh on page 256