Network and Remote Control Operation
R&S
®
SGS100A
180User Manual 1173.9105.02 ─ 12
On the R&S SGS the LXI functionality is already installed and enabled. Thus, the
instrument can be accessed via any web browser (like the Microsoft Internet Explorer)
to perform the following tasks:
Modifying network configurations
Remote control of the instrument
Performing SCPI remote diagnostics
10.5.1 Default State of the Network Settings
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter Value
TCP/IP mode DHCP + Auto IP Address
Dynamic DNS Enabled
ICMP ping Enabled
Password for LAN configuration LxiWebIfc
The LAN reset also resets the following parameters for the R&S SGS:
Parameter Value
Hostname <Instrument-specific host name>
Description Signal generator
Negotiation Auto detect
VXI-11 discovery Enabled
The LAN settings are configured using the instrument's LXI Browser Settings.
10.5.2 LXI Browser Settings
To access the instrument via the web browser:
Type in the instrument's host name or IP address in the address field of the
browser on your PC, for example "http://10.111.0.125".
Note: Do not add the missing zeros in the IP address, while opening the instru-
ment home page.