Remote Control Commands
R&S
®
SGS100A
206User Manual 1173.9105.02 ─ 12
Return values:
<All> 0 | 1 | OFF | ON
Example: See Chapter 11.1.1, "Performing General Task for Instrument
Setup", on page 191.
Usage: Query only
Manual operation: See "Adjust All" on page 129
:CALibration:FREQuency[:MEASure]?
Performs all adjustments which affect the frequency.
Return values:
<Synthesis> 0 | 1 | OFF | ON
Example: See Chapter 11.1.1, "Performing General Task for Instrument
Setup", on page 191.
Usage: Query only
Manual operation: See "Synthesis" on page 129
:CALibration:IQModulator:FULL?
Starts the adjustment of the I/Q modulator for the entire frequency range. The I/Q mod-
ulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Return values:
<Modulator> 0 | 1
Example: See Chapter 11.1.1, "Performing General Task for Instrument
Setup", on page 191.
Usage: Query only
Manual operation: See "I/Q Modulator" on page 130
:CALibration:IQModulator:LOCal?
Starts the adjustment of the I/Q modulator for the current frequency. The I/Q modulator
is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Example:
See Chapter 11.1.1, "Performing General Task for Instrument
Setup", on page 191.
Usage: Query only
Manual operation: See "Adjust I/Q Modulator at Current Frequency" on page 129
:CALibration:LEVel[:MEASure]?
Starts all adjustments which affect the level.
CALibration Subsystem