Remote Control Commands
R&S
®
SGS100A
211User Manual 1173.9105.02 ─ 12
Parameters:
<Channel> <Channellnstr#1>[,<Channellnstr#2>,...]
Returns a list of the used remote channels of the available
extensions, one interface per instrument. If an extension uses
more than one remote channels, the fastest one is returned.
Example: See Chapter 11.1.5, "Enabling and Configuring an Extension
Mode", on page 197.
Usage: Query only
Manual operation: See "Available Instruments" on page 126
:EXTension:INSTruments:REMote:LAN:NAME?
Queries the IP addresses/instrument names of the available extension instruments.
Return values:
<IPAddress> <IPAdr_Inst#1>[,<IPAdr_Inst#2>,...]
Example: See Chapter 11.1.5, "Enabling and Configuring an Extension
Mode", on page 197.
Usage: Query only
Manual operation: See "Available Instruments" on page 126
:EXTension:INSTruments:REMote:SERial?
Queiries the serial numbers of the available extension instruments.
Parameters:
<Serial> <SerialNumberInstr#1>[, <SerialNumberInstr#2>,...]
Example: See Chapter 11.1.5, "Enabling and Configuring an Extension
Mode", on page 197.
Usage: Query only
Manual operation: See "Available Instruments" on page 126
:EXTension:BUSY[:STATe]?
Queries the state of the extension instrument.
Return values:
<State> 0 | 1 | OFF | ON
0
The extension is connected to the master instrument and can be
remotely operated.
1
The extension is busy, i.e. in standby or locked state, or is per-
forming a time consuming operation.