Remote Control Commands
R&S
®
SGS100A
226User Manual 1173.9105.02 ─ 12
:SENSe<ch>[:POWer]:LOGGing:STATe............................................................................233
:SENSe<ch>[:POWer]:OFFSet........................................................................................233
:SENSe<ch>[:POWer]:OFFSet:STATe..............................................................................233
:SENSe<ch>[:POWer]:SNUMber?................................................................................... 233
:SENSe<ch>[:POWer]:SOURce.......................................................................................234
:SENSe<ch>[:POWer]:STATus[:DEVice]?.........................................................................234
:SENSe<ch>[:POWer]:SVERsion?...................................................................................234
:SENSe<ch>[:POWer]:TYPE?.........................................................................................235
:SENSe<ch>[:POWer]:ZERO.......................................................................................... 235
:SLISt[:LIST]?
Returns a list of all detected sensors in a comma-separated string.
Return values:
<SensorList> String of comma-separated entries
Each entry contains information on the sensor type, serial num-
ber and interface.
The order of the entries does not correspond to the order the
sensors are displayed in the "NRP Sensor Mapping" dialog.
Example: See Example "Detecting and assigning a power sensor"
on page 224.
Usage: Query only
Manual operation: See " Sensor Mapping List " on page 104
:SLISt:SCAN[:STATe] <State>
Starts the search for R&S NRP power sensors, connected in the LAN or via the
USBTMC protocol.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Detecting and assigning a power sensor"
on page 224.
Manual operation: See " Scan " on page 105
:SLISt:ELEMent<ch>:MAPPing <Mapping>
Assigns an entry from the :SLISt[:LIST]? to one of the four sensor channels.
Parameters:
<Mapping> SENS1 | SENSor1 | SENS2 | SENSor2 | SENS3 | SENSor3 |
SENS4 | SENSor4 | UNMapped
Sensor channel.
*RST: UNMapped