Remote Control Commands
R&S
®
SGS100A
227User Manual 1173.9105.02 ─ 12
Example: See Example "Detecting and assigning a power sensor"
on page 224.
Manual operation: See " Sensor Mapping List " on page 104
:SLISt:SENSor:MAP <SensorId>, <Mapping>
Assigns a sensor directly to one of the sensor channels, using the sensor name and
serial number.
To find out the the sensor name and ID, you can get it from the label of the R&S NRP,
or using the command :SLISt:SCAN[:STATe]. This command detects all R&S NRP
power sensors connected in the LAN or via 'USBTMC protocol.
Setting parameters:
<SensorId> string
<Mapping> enum
Example: See Example "Detecting and assigning a power sensor"
on page 224.
Usage: Setting only
Manual operation: See " Sensor Mapping List " on page 104
:INITiate<hw>[:POWer]:CONTinuous <Continuous>
Switches the local state of the continuous power measurement by R&S NRP power
sensors on and off. Switching off local state enhances the measurement performance
during remote control.
The remote measurement is triggered with :READ<ch>[:POWer]?). This command
also returns the measurement results. The local state is not affected, measurement
results can be retrieved with local state on or off.
Parameters:
<Continuous> 0 | 1 | OFF | ON
*RST: 0
Example:
INIT1:CONT ON
Switches local state of continuous power measurement on.
Manual operation: See " State " on page 109
:READ<ch>[:POWer]?
Triggers power measurement and displays the results.
Note: This command does not affect the local state, i.e. you can get results with local
state on or off. For long measurement times, we recommend that you use an SRQ for
command synchronization (MAV bit).
Suffix:
<ch>
.
1 to 3
SENSe, READ, INITiate and SLISt Subsystems