Remote Control Commands
R&S
®
SGS100A
236User Manual 1173.9105.02 ─ 12
BBBYpass
Baseband bypass mode
*RST: NORMal
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Manual operation: See "Operation Mode" on page 80
[:SOURce]:FREQuency[:CW|FIXed] <Cw>
Sets the RF frequency at the RF output connector of the selected instrument.
Note: Enabled frequency offset affects the result of this query. The query returns the
frequency, including frequency offset.
See [:SOURce]:FREQuency:OFFSet on page 236.
Parameters:
<Cw> float
*RST: 1 GHz
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Manual operation: See "Freq/Freq (Offs)" on page 66
[:SOURce]:FREQuency:OFFSet <Offset>
Sets a frequency offset, for example include the frequency shift of downstream instru-
ment.
Note: Enabled frequency offset affects the result of the query :SOURce:FRE
Quency:CW?
The query returns the frequency, including frequency offset.
Parameters:
<Offset> float
Range: -3e9 to 3e9
Increment: 0.001
*RST: 0
Example:
SOURce:FREQuency:OFFSet 0
SOURce:FREQuency:CW 6000000000
SOURce:FREQuency:OFFSet 20000000
SOURce:FREQuency:CW?
// 6020000000
Manual operation: See "Offset" on page 81
[:SOURce]:LOSCillator:SOURce <Source>
Selects the source of the local oscillator signal.
SOURce Subsystem