Remote Control Commands
R&S
®
SGS100A
248User Manual 1173.9105.02 ─ 12
Example: see Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193
Manual operation: See "Crest Factor" on page 121
[:SOURce]:IQ:WBSTate <State>
Selects optimized settings for wideband modulation signals.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Manual operation: See "I/Q Wideband" on page 121
11.16
SOURce:PHASe Subsystem
[:SOURce]:PHASe......................................................................................................... 248
[:SOURce]:PHASe:REFerence........................................................................................248
[:SOURce]:PHASe <Phase>
Specifies the phase variation relative to the current phase.
Parameters:
<Phase> float
Range: -360 to 360
Increment: 0.1
*RST: 0
Default unit: DEG
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Manual operation: See "Delta Phase" on page 82
[:SOURce]:PHASe:REFerence
Adopts the phase set with command [:SOURce]:PHASe as the current phase.
Example:
See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Usage: Event
Manual operation: See "Reset Delta Phase Display" on page 82
SOURce:PHASe Subsystem