Remote Control Commands
R&S
®
SGS100A
257User Manual 1173.9105.02 ─ 12
[:SOURce<hw>]:PULM:POLarity <Polarity>
Sets the polarity of the pulse modulator signal. This command is effective only for an
external modulation signal.
Parameters:
<Polarity> NORMal | INVerted
NORMal
The RF signal is suppressed during the pulse pause.
INVerted
The RF signal is suppressed during the pulse.
*RST: NORMal
Example:
PULM:SOUR EXT
selects the external modulation source.
Example:
PULM:POL INV
selects inverted polarity.
Manual operation: See "Polarity" on page 114
[:SOURce<hw>]:PULM:SOURce <Source>
Selects the source for pulse modulation.
Parameters:
<Source> INTernal | EXTernal
INTernal
The internal pulse generator is used for the pulse modulation.
EXTernal
The signal applied externally via the trigger connector is used for
the pulse modulation.
*RST: INTernal
Manual operation: See "Source" on page 114
[:SOURce<hw>]:PULM:STATe <State>
Activates the pulse modulation.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
PULM:STAT ON
activates pulse modulation.
Manual operation: See "State" on page 114
SOURce:PULM Subsystem