Remote Control Commands
R&S
®
SGS100A
261User Manual 1173.9105.02 ─ 12
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
ROSC:SOUR INT
Selects the internal source.
ROSC:ADJ ON
Activates use of a user-defined adjustment value.
ROSC:ADJ:VAL 1400
Sets the adjustment value to 1400.
Manual operation: See "Adjustment Active" on page 87
[:SOURce]:ROSCillator:EXTernal:SBANdwidth
<SBandwidth>
Sets the synchronization bandwidth for an external reference signal.
Parameters:
<SBandwidth> WIDE | NARRow
NARRow
The synchronization bandwidth is 40 Hz.
WIDE
Synchronization bandwidth is 250 Hz.
*RST: n.a. (factory preset)
Example:
ROSC:SOUR EXT
Selects the external source.
ROSC:EXT:FREQ 10 MHz
Informs the instrument that the external reference has a fre-
quency of 10 MHz.
ROSC:EXT:SBAN WID
Selects wideband setting for synchronization bandwidth.
Manual operation: See "Synchronization Bandwidth" on page 85
11.20 STATus Subsystem
This system contains the commands for the status reporting system. See also Chap-
ter A.5, "Status Reporting System", on page 302 for detailed information.
*RST on page 202 has no effect on the status registers.
Value ranges
Queries return the current value of the respective register, which permits a check of
the device status.
Return values: A decimal value in the range 0 to 32767 (=2
15
-1)
The configuration commands set the respective register thus determining which
status changes of the R&S SGS cause the status registers to be changed.