Instrument Tour
R&S
®
SGS100A
32User Manual 1173.9105.02 ─ 12
The [RF ON] key switches the RF signal on or off. If activated, the button is green.
[REF EXT]
The [REF EXT] LED indicates the status of the external reference.
●
Green indicates that the instrument can synchronize to the external clock.
●
Red indicates that the instrument cannot synchronize to the external clock.
●
No light indicates that the internal reference is used.
[ERROR / WARNING]
The [ERROR / WARNING] LED indicates the status of the R&S SGS.
●
Red indicates that an error has occurred, e.g. temperature exceeded or power fail-
ure.
●
Blinking orange indicates a running process (e.g calibration, self test).
●
No light indicates that no errors or warnings have occurred.
For details on errors or warnings, refer to the graphical user interface on a remote PC
("SGMA-GUI > Info").
For more information, refer to the "Error Messages" section in the user manual.
[LAN]
The [LAN] key indicates if a LAN connection is established.
Green LED indicates that a network cable is connected to the instrument and no error
in the DHCP connection is detected. But it does not indicate that the remote control
connection is established for all cases. For example, if the instrument is set to use a
"Static" IP address the instrument may still not be visible in the network.
Pressing the key resets the LAN interface settings, i.e the "IP Adress Mode" is reset to
DHCP.
[ID]
Pressing the [ID] key while the instrument is active, opens the "SGMA-GUI > Setup >
Instruments > Configure Instruments > Edit Instrument" dialog of the corresponding
instrument on the remote controller.
Front Panel Tour