Understanding the R&S SGMA-GUI Software
R&S
®
SGS100A
67User Manual 1173.9105.02 ─ 12
Note: The SCPI command [:SOURce]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display, that means the level containing off-
set while [:SOURce]:POWer:POWer sets the level at the RF output connector.
Remote command:
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude] on page 251
[:SOURce]:POWer:POWer on page 250
Ref. Oscillator Source/Ext Ref On/Off
Determines whether the internal built-in oscillator (TXCO or OXCO) is used as a refer-
ence source or if an external reference is used. The internal reference oscillator OCXO
requires the additional option R&S SGS-B1.
To feed in an external instrument reference, use the input connector [REF/LO IN]. To
output the reference frequency at the output [REF/LO OUT], select "SGMA-GUI >
Instrument Name > Ref. Oscillator > REF/LO Output > REF".
See also Chapter 9.4, "How to Configure the Reference Oscillator Source",
on page 152.
"Int"
The internal reference signal of 10 MHz is used.
"Ext"
An external reference signal is used. The frequency of the external
reference signal must be selected with the parameter "SGMA-GUI >
Instrument Name > Ref. Oscillator > Ex. Ref. Input Frequency".
Remote command:
[:SOURce]:ROSCillator:SOURce on page 260
LO Scr Ext
This icon indicates that the internal local oscillator is switched off.
See "Source" on page 83.
6.4 Working with R&S SGMA-GUI
This section explains how to work with the R&S SGMA-GUI software and perform con-
figuration tasks for manual operation of the instruments.
6.4.1 Storing and Loading Settings
To proceed work with a particular configuration of the instruments in the R&S SGMA-
GUI, it is useful to save the used settings and load them again later.
How to store and load settings
1. Select "SGMA-GUI main panel > File > Save As".
2. Navigate to the desired directory and enter the filename.
The extension *.savrcl is applied automatically.
The current settings of the software are saved to the selected file.