Understanding the R&S SGMA-GUI Software
R&S
®
SGS100A
76User Manual 1173.9105.02 ─ 12
:INSTruments:GPIB:BOARd..............................................................................................78
:INSTruments:SCAN........................................................................................................ 78
:INSTruments:SCAN:HNPRefix......................................................................................... 78
:INSTruments:SERial........................................................................................................78
:INSTruments:TYPE......................................................................................................... 79
:INSTruments:ACTive[:STATe] <State>
Enables/disables the instrument for the R&S SGMA-GUI. The main panel of this soft-
ware displays only activated instruments.
Parameters:
<State> List of BOOL-values
<StateInstr#1>,<StateInstr#2>,...
0, 1, ON, OFF
Example: See Chapter 6.5.1, "Configuring Instruments in the R&S SGMA-
GUI", on page 74.
Manual operation: See "Available Instruments" on page 55
:INSTruments:COUNt?
Queries the number of the currently available instruments.
Return values:
<Count> float
Range: 0 to 12
Example: See Chapter 6.5.1, "Configuring Instruments in the R&S SGMA-
GUI", on page 74.
Usage: Query only
Manual operation: See "Available Instruments" on page 55
:INSTruments:EACCess[:STATe] <State>
"Locks" the instruments, meaning the instrument is reserved and can be operated
exclusively from the remote PC that sent this SCPI command.
Tip: It is recommended to lock the instrument prior to further configuration.
Parameters:
<State> List of BOOL-values
<LockInstr#1>,<LockInstr#2>,...
0,1,OFF,ON
Example: See Chapter 6.5.1, "Configuring Instruments in the R&S SGMA-
GUI", on page 74.
Manual operation: See "Exclusive Access" on page 58
Remote Control of the R&S SGMA-GUI