EasyManua.ls Logo

Rohde & Schwarz SGS100A - Reference Oscillator

Rohde & Schwarz SGS100A
324 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Generator Settings
R&S
®
SGS100A
84User Manual 1173.9105.02 ─ 12
"Ext"
The signal fed-in at the [REF/LO IN] input connector is used as signal
source.
An icon in the block diagram indicates that an external LO source is
used.
Note: The local oscillator input/output requires the additional software
option R&S SGS-K90.
Remote command:
[:SOURce]:LOSCillator:SOURce on page 236
REF/LO Output
Determines the signal provided at the output connector [REF/LO OUT].
See also Chapter 9.6, "How to Define the Signal at the REF/LO OUT Connector",
on page 154.
"OFF"
No signal is provided.
"LO"
The signal of the local oscillator (LO) is available at the [REF/LO
OUT] connector.
"REF"
The signal of the reference oscillator is available at the [REF/LO
OUT] connector.
Remote command:
:CONNector:REFLo:OUTPut on page 208
7.4 Reference Oscillator
The R&S SGS is equipped with an internal TXCO reference oscillator. If equipped with
the additional HW option R&S SGS-B1, an OXCO can be used as an internal refer-
ence frequency source for the synthesizer. Moreover, an external reference signal
source can be connected to the REF/LO IN connector of the instrument.
Refer to Figure 5-9 for visualization of the signal flow.
To access this dialog, select "SGMA-GUI > Instrument Name > Reference Oscilla-
tor".
Reference Oscillator

Table of Contents