Improving Level Performance
R&S
®
SMA100B
243User Manual 1178.3834.02 ─ 05
State - Pulse Data Analysis
Enables pulse data analysis. The measurement is started with the "Start Cont." button
in the main measurement diagram.
Remote command:
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:STATe on page 461
Base - Pulse Data Analysis
Selects how the threshold parameters are calculated, either voltage related or power
related. The voltage-related parameters represent the normal case, as the usual repre-
sentation when defining the pulse parameters (rise/fall time, pulse width) is U(t). To
achieve a display with equivalent power-related values, the voltage-related threshold
values must be converted (squared) (see example in table below).
Distal Mesial Proximal
Voltage related: 90% 50% 10%
Power related: 81% 25% 1 %
log. Scale (for example):
(approximately, difference between top- base
power > 30 dB )
-0.9dB -6dB -20dB
Remote command:
:SENSe<ch>[:POWer]:SWEep:TIME[:SENSor]:PULSe:THReshold:BASE
on page 461
:TRACe<ch>[:POWer]:SWEep:PULSe:THReshold:BASE? on page 494