Improving Level Performance
R&S
®
SMA100B
246User Manual 1178.3834.02 ─ 05
Duty Cycle = (pulse duration / pulse period) * 100
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DCYCle? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DCYCle:DISPlay:
ANNotation[:STATe] on page 493
Pulse Width ← Notifications - Pulse Data Analysis
Indicates the pulse duration of the pulse data measurement in seconds.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DURation? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DURation:DISPlay:
ANNotation[:STATe] on page 493
Pulse Period ← Notifications - Pulse Data Analysis
Indicates the time the pulse signal needs to complete one cycle.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod:DISPlay:
ANNotation[:STATe] on page 493
Pulse Off Time ← Notifications - Pulse Data Analysis
Determines the time the pulse signal is low, that means as long as the signal level is
below the proximal value.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:SEParation? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod:DISPlay:
ANNotation[:STATe] on page 493
Transition Times - Pulse Data Analysis
Selects the transition parameters of the pulse signal to be indicated below the mea-
surement diagram.
The R&S NRP-Z sensor searches for the first rising edge and the first falling edge in
the trace.
Rise Time / Fall Time ← Transition Times - Pulse Data Analysis
Displays the time the signal requires from crossing low reference until it reaches high
reference level and vice versa.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:NEGative:
DURation? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:NEGative:
DURation:DISPlay:ANNotation[:STATe] on page 493
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:POSitive:
DURation? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:POSitive:
DURation:DISPlay:ANNotation[:STATe] on page 493
Using Power Sensors