Improving Level Performance
R&S
®
SMA100B
247User Manual 1178.3834.02 ─ 05
Pulse Start Time / Pulse Stop Time ← Transition Times - Pulse Data Analysis
Displays the time when the pulse signal crosses the medial reference level.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:NEGative:
OCCurrence? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:NEGative:
OCCurrence:DISPlay:ANNotation[:STATe] on page 493
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:POSitive:
OCCurrence? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:POSitive:
OCCurrence:DISPlay:ANNotation[:STATe] on page 493
Overshoot (Rising Edge / Falling Edge) ← Transition Times - Pulse Data Analysis
Display the maximum value of the pulse signal following a rising transition and the min-
imum value of the signal after a falling transition, respectively.
Overshoot values are given in per cent of the pulse amplitude as shown below:
Overshoot(pos) = 100* (maximum - top level ) / (top level - base level)
Overshoot(neg) = 100* (base level - minimum) / (top level - base level)
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:POSitive:
OVERshoot? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:POSitive:
OVERshoot:DISPlay:ANNotation[:STATe] on page 493
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:NEGative:
OVERshoot? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:TRANsition:NEGative:
OVERshoot:DISPlay:ANNotation[:STATe] on page 493
Signal Power - Pulse Data Analysis
Selects the power parameters of the pulse signal to be indicated below the measure-
ment diagram.
Minimal / Peak / Average Power ← Signal Power - Pulse Data Analysis
Display the minimum, the maximum and the average power of the pulse signal in dBm.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:MINimum? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:MINimum:DISPlay:
ANNotation[:STATe] on page 493
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:MAXimum? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:MAXimum:DISPlay:
ANNotation[:STATe] on page 493
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:AVERage? on page 492
:TRACe<ch>[:POWer]:SWEep:MEASurement:POWer:AVERage:DISPlay:
ANNotation[:STATe] on page 493