Remote Control Commands
R&S
®
SMA100B
427User Manual 1178.3834.02 ─ 05
UP|DOWN
Varies the frequency step by step.
The frequency is increased or decreased by the value set with
the command :CSYNthesis:FREQuency:STEP.
Range: 100E3 to 1.5E9
Increment: 0.001
*RST: 10E6
Example: See Chapter 13.7, "CSYNthesis Subsystem", on page 425.
Manual operation: See " Frequency " on page 280
:CSYNthesis:POWer <Power>
Sets the power level of the generated clock signal.
Parameters:
<Power> float
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command :CSYNthesis:POWer:STEP[:INCRement].
Range: -24 to 10
Increment: 0.01
*RST: -20
Example: See Chapter 13.7, "CSYNthesis Subsystem", on page 425.
Manual operation: See " Level " on page 280
:CSYNthesis:OFFSet:STATe <State>
Activates a DC offset.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Chapter 13.7, "CSYNthesis Subsystem", on page 425.
Manual operation: See " DC Offset State " on page 280
:CSYNthesis:OFFSet <Offset>
Sets the value of the DC offset.
CSYNthesis Subsystem