Remote Control Commands
R&S
®
SMA100B
429User Manual 1178.3834.02 ─ 05
:CSYNthesis:POWer:STEP:MODE <Mode>
:CSYNthesis:FREQuency:STEP:MODE <Mode>
Defines the type of step size to vary the frequency and level at discrete steps.
Parameters:
<Mode> DECimal | USER
DECimal
Increases or decreases the level in steps of 10.
USER
Increases or decreases the value in increments, set with the
command:
:CSYNthesis:FREQuency:STEP
:CSYNthesis:POWer:STEP[:INCRement]
*RST: DECimal
Example: See Chapter 13.7, "CSYNthesis Subsystem", on page 425.
Manual operation: See "Variation Active" on page 281
:CSYNthesis:POWer:STEP[:INCRement] <Increment>
Sets the step width of the rotary knob and, in user-defined step mode, increases or
decreases the level.
Parameters:
<Increment> float
Range: 0 to 35
Increment: 0.01
*RST: 1
Example: See Chapter 13.7, "CSYNthesis Subsystem", on page 425.
Manual operation: See "Variation Step" on page 281
:CSYNthesis:FREQuency:STEP <Step>
Sets the step width of the rotary knob and, in user-defined step mode, increases or
decreases the frequency.
Parameters:
<Step> float
Range: 0 to 14999E5
Increment: 0.001
*RST: 1E6
Example: See Chapter 13.7, "CSYNthesis Subsystem", on page 425.
Manual operation: See "Variation Step" on page 281