Remote Control Commands
R&S
®
SMA100B
458User Manual 1178.3834.02 ─ 05
Options: Option R&S SMAB-K28
Manual operation: See " Level Offset State- Power Analysis " on page 239
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STARt <Start>
Sets the start frequency for the frequency power analysis with separate frequencies.
Parameters:
<Start> integer
Range: 0 to 1E12
*RST: 1E6
Example:
SENS2:SWE:FREQ:SENS:SRAN:STAT ON
Activates use of a separate frequency range for frequency ver-
sus power measurement for sensor 2.
SENS2:SWE:FREQ:SENS:STAR 2.0GHZ
Sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
Manual operation: See " Min Frequency - Power Analysis " on page 240
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STOP <Stop>
Sets the stop frequency for the frequency power analysis with separate frequencies.
Parameters:
<Stop> integer
Range: 0 to 1E12
*RST: 10E6
Example:
SENS:SWE:FREQ:SENS2:SRAN:STAT ON
Activates use of a separate frequency range for frequency ver-
sus power measurement.
SENS:SWE:FREQ:SENS2:STAR 2.0GHZ
Sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
SENS:SWE:FREQ:SENS2:STOP 2.9GHZ
Sets a sweep stop at 2.9 GHz irrespective of the current signal
generator frequency settings.
Manual operation: See " Max Frequency - Power Analysis " on page 241
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge[:STATe] <State>
Activates the use of a frequency range for the power measurement that is different to
the set signal generator frequency range. The separate frequency range is entered
with commands :SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:
STARt on page 458 and :SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:
SRANge:STOP on page 458.
Power Sensor Measurement Subsystems