Remote Control Commands
R&S
®
SMA100B
480User Manual 1178.3834.02 ─ 05
Example:
SENS:SWE:POW:RMOD SING
Selects single measurement.
Manual operation: See " Execution - Power Analysis " on page 234
:SENSe[:POWer]:SWEep:POWer:SPACing[:MODE] <Mode>
Selects the spacing for the frequency power analysis.
Parameters:
<Mode> LINear
*RST: LINear
Example:
SENS:SWE:FREQ:SPAC:MODE LIN
Sets linear spacing of the sweep.
Manual operation: See " Spacing - Power Analysis " on page 233
:SENSe[:POWer]:SWEep:POWer:STARt <Start>
Sets the start level for the power versus power measurement.
Parameters:
<Start> float
Range: -145 to 20
Increment: 0.01
*RST: 1MHZ
Example:
SENS:SWE:POW:STAR -20DBM
Sets the start level to -20 dBm
Manual operation: See "Min - Power Analysis" on page 231
:SENSe[:POWer]:SWEep:POWer:STEPs <Steps>
Sets the number of measurement steps for the power versus power measurement.
Parameters:
<Steps> integer
Range: 10 to 1000
*RST: 500
Example:
SENS:SWE:POW:STEP 500
Sets the 500 measurement steps.
Manual operation: See " Steps - Power Analysis " on page 232
:SENSe[:POWer]:SWEep:POWer:STOP <Stop>
Sets the stop level for the power versus power measurement.
Power Sensor Measurement Subsystems