Remote Control Commands
R&S
®
SMA100B
481User Manual 1178.3834.02 ─ 05
Parameters:
<Stop> float
Range: -145 to 20
Increment: 0.01
*RST: 40
Example:
SENS:SWE:POW:STOP 20.0DBM
Sets the stop level to 20 dBm.
:SENSe[:POWer]:SWEep:POWer:TIMing[:MODE] <Mode>
Selects the timing mode of the measurement.
Parameters:
<Mode> FAST | NORMal | HPRecision
FAST
Selection FAST leads to a fast measurement with a short inte-
gration times for each measurement step.
NORMal
NORMal leads to a longer but more precise measurement due
to a higher integration time for each step.
*RST: NORMal
Example:
SENS:SWE:POW:TIM:MODE FAST
Selects fast mode.
Manual operation: See " Timing - Power Analysis " on page 233
:SENSe[:POWer]:SWEep:POWer:YSCale:AUTO <Auto>
Activates autoscaling of the Y axis of the diagram.
Parameters:
<Auto> OFF | CEXPanding | FEXPanding | CFLoating | FFLoating
OFF
Auto scaling is deactivated. When switching from activated to
deactivated Auto scaling, the scaling is maintained. When
switching from deactivated to activated Auto scaling, the scaling
is reset to min = max = 0.
CEXPanding | FEXPanding
Auto scale is activated. The scaling of the Y-axis is selected in
such a way, that the trace is always visible. To this end, the
range is expanded if the minimum or maximum values of the
trace move outside the current scale. The step width is 5 dB for
selection course and variable in the range of 0.2 db to 5 dB for
selection fine.
Power Sensor Measurement Subsystems