Remote Control Commands
R&S
®
SMA100B
483User Manual 1178.3834.02 ─ 05
Parameters:
<Minimum> float
Range: -200 to 100
Increment: 0.01
*RST: -40
Example:
SENS:SWE:POW:YSC:MIN -10DBM
Sets -10 dBm as the lower limit of the measurement diagram.
Manual operation: See " Min - Max Level - Power Analysis " on page 235
:SENSe[:POWer]:SWEep:RMODe
<RMode>
Selects single or continuous mode for power analysis (all measurement modes).
Parameters:
<RMode> SINGle | CONTinuous
*RST: SINGle
Example:
SENS:SWE:RMOD SING
Selects single measurement.
Manual operation: See " Execution - Power Analysis " on page 234
:SENSe[:POWer]:SWEep:TIME:AVERage[:COUNt] <Count>
Selects the averaging factor in time mode. The count number determines how many
measurement cycles are used to form a measurement result. Higher averaging counts
reduce noise but increase the measurement time. Averaging requires a stable trigger
event so that the measurement cycles have the same timing.
Parameters:
<Count> 1 | 2 | 4 | 8 | 16 | 32 | 64 | 128 | 256 | 512 | 1024
*RST: 1
Example:
SENS:SWE:MODE TIME
Selects time mode.
SENS:SWE:TIME:AVER 128
Selects averaging factor 128.
Options: Option R&S SMAB-K28
Manual operation: See " Average - Power Analysis " on page 233
:SENSe[:POWer]:SWEep:TIME:REFerence:DATA:COPY
Generates a reference curve for "Time" measurement.
Example:
SENS:POW:SWE:TIME:REF:DATA:COPY
Generates a reference curve in time mode.
Usage: Event
Manual operation: See "Save To Ref" on page 228
Power Sensor Measurement Subsystems