:SENSe[:POWer]:SWEep:TIME:REFerence:DATA:POINts?
Queries the number of points from the reference curve in "Time" measurement.
Return values:
<Points> integer
Range: 10 to 1000
*RST: 0
Example:
SENS:POW:SWE:TIME:REF:DATA:POIN?
Queries the number of points from the reference curve in time
mode.
Usage: Query only
:SENSe[:POWer]:SWEep:TIME:REFerence:DATA:XVALues <XValues>
Sets or queries the x values of the two reference points, i.e. "Time X (Point A)" and
"Time X (Point B) "in "Time" measurement.
Parameters:
<XValues> string
Example:
SENSe:POW:SWE:TIME:REF:DATA:XVAL 5,45
Sets the x value of reference "Point A" to 5 μs, and the value of
"Point B" to 45 μs.
Manual operation: See "Time (X) / Pow (Y)" on page 228
:SENSe[:POWer]:SWEep:TIME:REFerence:DATA:YVALues <YValues>
Sets or queries the y values of the two reference points, i.e. "Power Y (Point A)" and
"Power Y (Point B)" in "Time" measurement.
Parameters:
<YValues> string
Example:
SENSe:POW:SWE:TIME:REF:DATA:YVAL -30,10
sets the y value of reference "Point A" to -30 dBm, and the value
of "Point B" to 10 dBm.
Manual operation: See "Time (X) / Pow (Y)" on page 228
:SENSe[:POWer]:SWEep:TIME:RMODe <RMode>
Selects single or continuous mode for measurement mode time in power analysis.
Parameters:
<RMode> SINGle | CONTinuous
*RST: CONTinuous
Example:
SENS:SWE:TIME:RMOD SING
Selects single measurement.