Remote Control Commands
R&S
®
SMA100B
503User Manual 1178.3834.02 ─ 05
Note: This command does not affect the local state, i.e. you can get results with local
state on or off. For long measurement times, we recommend that you use an SRQ for
command synchronization (MAV bit).
Suffix:
<ch>
.
1 to 3
Return values:
<Power> float or float,float
The sensor returns the result in the unit set with command :
SENSe<ch>:UNIT[:POWer]
Certain power sensors, such as the R&S NRP-Z81, return two
values, first the value of the average level and - separated by a
comma - the peak value.
Example:
SENS1:UNIT DBM
// Selects unit dBm for presentation of measurement result
READ1?
// Queries the measurement result of the sensor
// -45.6246576745440230
// -45.6 dBm were measured at the given frequency
Example: R&S NRP-Z81
READ1?
-55.62403263352178,-22.419472478812476
-55.6 dBm is the measured average level, -22. 4 dBm is the measured peak level at the given frequency
Usage: Query only
Manual operation: See " Level (Peak) / Level (Average) " on page 218
:SENSe<ch>:UNIT[:POWer] <Power>
Selects the unit (Watt, dBm or dBμV) of measurement result display, queried with :
READ<ch>[:POWer]?.
Parameters:
<Power> DBM | DBUV | WATT
*RST: DBM
Example:
SENS2:UNIT DBM
Selects dBm as unit for the measured value returned by com-
mand READ.
READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation: See " Level (Peak) / Level (Average) " on page 218
:SENSe<ch>[:POWer]:APERture:DEFault:STATe <UseDefAp>
Deactivates the default aperture time of the respective sensor.
SENSe, READ, INITiate and SLISt Subsystems