Remote Control Commands
R&S
®
SMA100B
611User Manual 1178.3834.02 ─ 05
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RUNNing?
Queries the current status of the LF frequency sweep mode.
Return values:
<State> 0 | 1 | OFF | ON
Example: See Example "Setup an LF sweep" on page 599.
Usage: Query only
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SHAPe
<Shape>
Sets the cycle mode for a sweep sequence (shape).
Parameters:
<Shape> SAWTooth | TRIangle
*RST: SAWTooth
Example: See Example "Setup an LF sweep" on page 599.
Manual operation: See " Shape" on page 173
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing <Spacing>
Selects linear or logarithmic sweep spacing.
Parameters:
<Spacing> LINear | LOGarithmic
*RST: LINear
Example: See Example "Setup an LF sweep" on page 599.
Manual operation: See " Spacing" on page 174
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
<Logarithmic>
Sets the step width factor for logarithmic sweeps to calculate the frequencies of the
steps.
For information on how the value is calculated and the interdependency with other
parameters, see Chapter 6.2.1, "Correlating Parameters in Sweep Mode", on page 164
Parameters:
<Logarithmic> float
The unit is mandatory
Range: 0.01 to 100
Increment: 0.01
*RST: 1
Default unit: PCT
Example: See Example "Setup an LF sweep" on page 599.