Remote Control Commands
R&S
®
SMA100B
659User Manual 1178.3834.02 ─ 05
:SYSTem:ERRor:GNEXt?
Similar to :SYSTem:ERRor[:NEXT]?, but queries the next entry from the global per-
sistant error/event queue.
Return values:
<NextGlobalError> string
Error/event number, "Error/event description> [;Device depend-
ent info]"
An error number and a short description of the error.
Positive error numbers are instrument-dependent. Negative
error numbers are reserved by the SCPI standard.
Example:
SYST:ERR:GNEX?
Returns the next error message from the global error queue.
Usage: Query only
:SYSTem:ERRor:HISTory:CLEar
Clears the error history.
Example:
See Example "Querying the error queue" on page 655
Usage: Event
Manual operation: See "Clear History" on page 706
:SYSTem:ERRor:STATic?
Returns a list of all errors existing at the time when the query is started. This list corre-
sponds to the display on the info page under manual control.
Return values:
<StaticErrors> string
Example: See Example "Querying the error queue" on page 655
Usage: Query only
Manual operation: See "Static Errors/Error History" on page 706
:SYSTem:ULOCk <Mode>
Locks or unlocks the user interface of the instrument.
Parameters:
<Mode> ENABled | DONLy | DISabled | TOFF | VNConly
ENABled
Unlocks the display, the touchscreen and all controls for the
manual operation.
SYSTem Subsystem