Remote Control Commands
R&S
®
SMA100B
662User Manual 1178.3834.02 ─ 05
:SYSTem:SECurity:VOLMode[:STATe] <SecPassWord>, <MmemProtState>
Activates volatile mode, so that no user data can be written to the internal memory per-
manently.
To enable volatile mode, reboot the instrument. Otherwise the change has no effect.
Parameters:
<MmemProtState> 0 | 1 | OFF | ON
*RST: 0
Setting parameters:
<SecPassWord> string
Current security password
The default password is 123456.
Example:
SYSTem:SECurity:VOLMode:STATe "123456", 1
SYSTem:REBoot
Manual operation: See "Volatile Mode" on page 336
:SYSTem:COMMunicate:GPIB:LTERminator <LTerminator>
Sets the terminator recognition for remote control via GPIB interface.
Parameters:
<LTerminator> STANdard | EOI
EOI
Recognizes an LF (Line Feed) as the terminator only when it is
sent with the line message EOI (End of Line). This setting is rec-
ommended particularly for binary block transmissions, as binary
blocks may coincidentally contain a characater with value LF
(Line Feed), although it is not determined as a terminator.
STANdard
Recognizes an LF (Line Feed) as the terminator regardless of
whether it is sent with or without EOI.
*RST: n.a. (factory preset: STANdard)
Example: See Example "Finding out the used VISA ressource strings"
on page 654.
:SYSTem:COMMunicate:GPIB:RESource?
Queries the visa resource string for remote control via the GPIB interface.
To change the GPIB address, use the command :SYSTem:COMMunicate:GPIB[:
SELF]:ADDRess.
Return values:
<Resource> string
Example: See Example "Finding out the used VISA ressource strings"
on page 654.
SYSTem Subsystem