RF Signal Configuration
R&S
®
SMA100B
65User Manual 1178.3834.02 ─ 05
Frequency
Sets the RF frequency.
This frequency is output at the RF connector. It does not consider an Offset or multipli-
cation factor (Multiplier).
See "RF frequency and level display with a downstream instrument" on page 63.
Note: Suppressed values in the status bar
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
The display has been disabled for security reasons.
See:
Annotation Frequency
Annotation Amplitude
The display is disabled when list mode is running, see Chapter 6, "List and Sweep
Mode", on page 154.
Remote command:
[:SOURce<hw>]:FREQuency[:CW|FIXed] on page 592
Main PLL Bandwidth
Selects the PLL (Phase Locked Loop) bandwidth of the main synthesizer.
"Normal"
Default main PLL bandwidth.
The instrument provides the maximum modulation bandwidth and
FM/PhiM deviation.
"Narrow"
Sets the narrow PLL bandwidth.
Remote command:
[:SOURce<hw>]:FREQuency:PLL:MODE on page 597
Offset
Sets the frequency offset.
This value represents the frequency shift of a downstream instrument, like for example
a mixer.
The "Frequency" value displayed in the status bar is the resulting frequency, as it is at
the output of the downstream instrument. The frequency at the R&S SMA100B RF out-
put is not changed.
See "RF frequency and level display with a downstream instrument" on page 63.
Remote command:
[:SOURce<hw>]:FREQuency:OFFSet on page 594
Multiplier
Sets the multiplication factor for the RF frequency.
RF Frequency Settings