Remote Control Commands
R&S
®
SMA100B
669User Manual 1178.3834.02 ─ 05
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:IRES "Test Device"
// Defines identification string 'test device'
*IDN?
// Response: 'test device'
Manual operation: See "IDN String" on page 364
:SYSTem:ORESponse <OResponse>
Defines the user defined response string for *OPT.
Note: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYST:ORES is discarded.
Parameters:
<OResponse> string
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:ORES "Test Option"
// Defines the OPT string 'test option'
*OPT?
// Response: 'test option'
Manual operation: See "OPT String" on page 364
:SYSTem:LANGuage <Language>
Sets the remote control command set.
Parameters:
<Language> string
Example:
SYSTem:LANGuage "SCPI"
// selects SCPI command set
Manual operation: See "Language" on page 363
:SYSTem:INFormation:SCPI <InfoString>
Inserts system information in recorded SCPI command lists, for example information
on a missing command.
Parameters:
<InfoString> string
Example:
SYST:INF:SCPI "missing command"
enters the information into a recorded SCPI command list.
SYSTem Subsystem