Troubleshooting and Error Messages
R&S
®
SMA100B
705User Manual 1178.3834.02 ─ 05
To display information on static errors and error history
1. In the taskbar, select the "Info" icon.
2. For some messages, the information line appears briefly on the home screen.
To open the dialog, in the "Info" line select the "Info" button.
The "Static Errors" dialog lists the last monitored messages chronologically and
displays additional information on the highlighted message.
The "Error History" dialog lists the accumulated messages with a short descrip-
tion.
Volatile errors are reported once, at the time they appear. If identical errors
occur subsequently, they are not reported repeatedly.
In the "Info" line, identical errors are displayed repeatedly only if the original
error has already disappeared from the display. If queried by SCPI command,
identical errors are only reported if the original error has already been retrieved
from (and hence not any more present in) the error queue.
Querying Error Messages