Remote Control Commands
R&S
®
SMA100B
423User Manual 1178.3834.02 ─ 05
:CALibration<hw>:ALL:TIME?......................................................................................... 424
:CALibration:DATA:FACTory:DATE?.................................................................................424
:CALibration<hw>:CONTinueonerror................................................................................425
:CALibration:ALL[:MEASure]? [<Force>]
Starts all internal adjustments that do not need external measuring equipment.
NOTICE: Risk of DUT damage
During level adjustments, the instrument temporarily applies high power at the RF out-
put. This high power can destroy a connected DUT.
Do not start level adjustments if DUT is connected. Disconnect the DUT and replace it
by a terminating resistor with adequate power rating. We recommend that you use a
50 Ohm, 10 W or larger terminating resistor.
Query parameters:
<Force> string
Return values:
<Measure> 0 | 1 | OFF | ON
Example:
CAL:ALL:MEAS?
0
Executes the adjustments of all instrument functions. When
completed, it indicates that the adjustment has been performed
successfully.
Usage: Query only
Manual operation: See "Adjust All" on page 695
:CALibration<hw>:ALL:DATE?
Queries the date of the most recently executed full adjustment.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Date> string
Example:
CAL:ALL:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Full Adjustment" on page 696
:CALibration<hw>:ALL:INFormation?
Queries the current state of the internal adjustment.
CALibration Subsystem